Appendix D
Recommended DW-MRI scan protocols

Date: May 21, 2010

Disclaimer: this document contains recommendations for optimal DW-MRI acquisition. We (Brain Innovation BV) are not responsible for any damages caused in any way resulting from following these recommendations.

Sequence
double-refocused single shot EPI reduces eddy current distortions.
Field of View
minimal 240 mm (square)
k-space coverage
symmetrical
Number of Slices
minimum needed to cover entire brain including cerebellum (usually > 55)
Slice orientation
axial
Voxel size
Isotropic, 2.5 x 2.5 x 2.5 minimum recommended
Matrix size
square 96x96, 128x128, etc
Slice gap
0 mm
Angulation
none, could cause mis-calculation of tensors
TR
shortest to accommodate all slices
TE
minimum needed for full acquisition
Zero-filling
No zero-filling or interpolation
Parallel imaging
SENSE (factor 2) or GRAPPA
Diffusion directions
> 6 directions for MD, > 20 directions for FA and fiber tracking, > 80 directions for HARDI applications
b0
1 b0 for every 6-10 diffusion weighted images
b-value
600-1200 s/mm2 for DTI, > 2000 for HARDI/q-ball etc
Averages
As many as possible given the scan time. Do not average real-time on the scanner, but save each repetition as a seperate data set.
Supported Raw data formats in BVQX
DICOM, DICOM MOSAIC, PAR/REC, Nifti via Nifticonverter plugin